Lab chiller is considered an instrument designed to provide the necessary conditions for experimentation and research, which can be moved on wheels, or small enough to be carried or set on a counter. Having the advantages of accuracy, sustainability, cost saving, convenience, safety, etc, CW-6200ANWTY chiller can be used to cool MRI machines, linear accelerators, CT scanners, radiation therapy equipment, etc.
TEYU water cooled lab chiller CW-6200ANSWTY does not need a fan to cool the condenser, which reduces noise and heat emission to the operating space, and is more green energy-saving. Using external circulating water to cooperate with the internal system for efficient refrigeration, smaller in size with a 6600W large cooling capacity with precise PID temperature control of ±0.5°C and less space occupation. Lab chiller CW-6200ANSWTY supports RS485 communication, and complaints with the CE, RoHS and REACH standards and comes with a 2-year warranty.
Model: CW-6200ANSWTY
Machine Size: 67X47X80cm (L X W X H)
Warranty: 2 years
Standard: CE, REACH and RoHS
Model | CW-6200ANSWTY |
Voltage | AC 1P 220-240V |
Frequency | 50Hz |
Current | 2.5~19.9A |
Max. power consumption | 3.52kW |
| 1.75kW |
2.38HP | |
| 22519Btu/h |
6.6kW | |
5674Kcal/h | |
Refrigerant | R-410a |
Precision | ±0.5℃ |
Reducer | Capillary |
Pump power | 0.37kW |
Tank capacity | 22L |
Inlet and outlet | Rp1/2"+ Rp 1/2" |
Max. pump pressure | 3.6bar |
Max. pump flow | 75L/min |
N.W. | 67Kg |
G.W. | 79Kg |
Dimension | 67X47X80cm (L X W X H) |
Package dimension | 73X57X105cm (L X W X H) |
The working current can be different under different working conditions. The above information is for reference only. Please subject to the actual delivered product.
* Cooling capacity: 6600W
* Active cooling
* Control accuracy: ±0.5°C
* Temperature control range: 5°C ~35°C
* Small size with large cooling capacity
*Stable working performance with low noise level and long lifespan
* High efficiency with low maintenance
* No heat interference to the operating room
Heater
Filter
US standard plug / EN standard plug
Digital temperature controller
The digital temperature controller offers high precision temperature control of ±0.5°C.
Dual water inlet and water outlet
Water inlets and water outlets are made from stainless steel to prevent potential corrosion or water leakage.
Modbus RS485 communication port integrated in the electrical connecting box
The RS485 communication port integrated in the electrical connecting box enables communication with the equipment to be cooled.